Fidji® Lumbar Cages
- Posterior, anterior, anterolateral and lateral cages
- PEEK-OPTIMA® cages
- Rounded tip and convex shape (0, 4°and 8°)
- Autostatic teeth
- Important bone graft area
PEEK
|
Poly-Ether-Ether-Ketone unique and unparalleled properties PEEK interbody fusion cages are more than simple spacers permitting vascularization between vertebral endplates; they mediate the application of physiological load conditions to the bone graft. This ensures continuity in the mechanical properties throughout the level involved, and consequently optimizes the conditions contributing to successful fusion. |
|
|
|
Mechanical Tests
| Posterior Cages (PLIF) | Anterior Cages (ALIF) | ||
| Static Compression Test | Yield Strength | 7.8kN | 18.9kN |
| Yield Displacement | 2.1mm | 1.2mm | |
| Strength to Failure | 10.1kN | 31.7kN | |
| Displacement to Failure | 3.55mm | 2.1mm | |
| Dynamic Compression Test (5 Hertz) | 5 million cycles under a load of 3.7kN | 5 million cycles under a load of 8.9kN |
Elasticity
- This oversized simulation illustrates the transient change in Fidji PLIF cages under loads of 50 kg.
|
|
|
|
Shape and Design
Angles
Shape
|
|
|
Radiopaque Markers
Primary Stability
|
|
|
Secondary Stability
|
|